Dual Energy X-ray for Quantitative Analysis of areal Bone Mineral Density (aBMD) using a stacked Flat Panel Detector (FPD)

Loading...
Thumbnail Image

Date

2025-04-23

Advisor

Karim, Karim Sallaudin

Journal Title

Journal ISSN

Volume Title

Publisher

University of Waterloo

Abstract

This study explores the application of a stacked flat panel detector (FPD) system for spectral dual-energy X-ray imaging aimed at quantifying areal bone mineral density (aBMD). The proposed technique enables single-exposure dual-energy acquisition using a conventional cone-beam X-ray source, eliminating the need for sequential exposures or specialized hardware. This method facilitates simultaneous high- and low-energy imaging by capturing energy-separated data directly at the detector level, thereby enhancing workflow efficiency and spatial alignment. Experimental validation using bone-equivalent phantoms demonstrates the system's potential for accurate aBMD estimation, with robustness against positional variation and tissue scatter. The findings suggest a promising pathway toward more accessible, portable, and cost-effective bone health assessment in clinical and point-of-care settings.

Description

Keywords

spectral x-ray, dual-energy x-ray, bone densitometry, medical physics, medical imaging